To investigate the possibility of a long-term applicable left ventricular assist device, a 23 mm outer diameter and 31 g weight implantable aortic valvo-pump was developed. It consists of a rotor and a stator; the rotor has a driven magnets assemble and an impeller, the stator has a motor coil with iron core and a outflow guide vane. The device locates the position of aortic valve, delivers the blood directly from left ventricle to aorta. Neither connecting conduits nor "bypass" circuits are necessary. Therefore, the device has promisingly better antithrombogenicity than other heart pumps. In hemodynamic testing, the pump can produce a blood flow of 7 l/min volume with 50 mmHg pressure increase at 15,000 rpm rotating speed, and at zero flow rate the pump can maintain a diastolic pressure over 80 mmHg at same rotating speed. For further studies the blood compatibility and the durability of the device are of most importance.
Equipment Failure Analysis, Heart Valve Prosthesis, Humans, Equipment Design, Heart-Assist Devices, Aorta
Equipment Failure Analysis, Heart Valve Prosthesis, Humans, Equipment Design, Heart-Assist Devices, Aorta
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 3 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |