Non-linear loading of the utilities supply introduced significant electromagnetic field (EMF) interference severe enough to disrupt electrocardiograph (ECG) monitoring and recordings in the new Emergency Department (ED) at the Royal Melbourne Hospital (RMH). This interference was evident even though standard Mu-metal shielding had been installed over the main hospital power feed which runs underneath the department. Investigations revealed that the source of the interference was due to 3rd harmonic currents flowing in the mains cable. This interference was suppressed by introducing third harmonic current into the main power cable in anti-phase to the interfering signal.
Equipment Failure Analysis, Electrocardiography, Electric Power Supplies, Electromagnetic Fields, Humans, Emergency Service, Hospital, Electronics, Medical
Equipment Failure Analysis, Electrocardiography, Electric Power Supplies, Electromagnetic Fields, Humans, Emergency Service, Hospital, Electronics, Medical
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |