doi: 10.1007/bf02477668
pmid: 5496688
Ultrafiltration rates (UFR) have been measured for different membranes in a modified dialysis test cell under a range of positive pressures from 50 to 300 Torr. The test cell interior is filled with 200 ml distilled water and water escape rate is measured through one 5·1 cm diameter disc. A linear relationship between UFR and applied pressure exists for cellulose materials and experimental results confirm those of previous workers. A new synthetic membrane presently under clinical evaluation at Strathclyde (Series 10 or “Permane”) shows much higher UFR for any applied pressure compared with cellulose films under the same conditions. A different water transport mechanism is evident from the resulting curves which are obtained by plotting “Permane” UFR against applied pressure. Several methods exist for controlling “Permane” UFR.
Methods, Pressure, Membranes, Artificial, Dialysis, Filtration, Permeability
Methods, Pressure, Membranes, Artificial, Dialysis, Filtration, Permeability
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 9 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |