doi: 10.1007/bf02474156
pmid: 5908038
The quoted life of pacemakers available commercially is from 2–5 yr. Experience with these units has shown that their life has not been more than 12–14 months and in some cases has been as low as 7 months. If assessment of the pacemaker's electrical condition is to be made after implant, any tests used must be based on the measurement of output characteristics. For any given unit these characteristics vary considerably with time, thus making assessment of the pacemaker's condition extremely difficult. The instability and relatively short life of these units has necessitated the design of an implantable unit with stable characteristics. The effects of shelf life, operating temperature, and pacemaker load impedance are considered and it is shown that the calculated and actual life are in agreement. A pacemaker design is described. This has proved to be stable and has a life of 2–3 yr. The circuit design principle is unique and mathematical analysis is used to show its advantages over conventional circuits.
Pacemaker, Artificial, Mathematics
Pacemaker, Artificial, Mathematics