A simple battery-powered constant-current pulse generator able to supply up to 200 μA at a maximum of 250 V is described. The tester incorporates a battery status indicator and an overload, incorrect operation and fault condition audible alarm. A simple test routine helps to ensure that the instrument is operating correctly before each application. For enhanced patient co-operation and comfort the tester is activated by the patient.
Electric Power Supplies, Dental Pulp Test, Humans, Electronics, Medical
Electric Power Supplies, Dental Pulp Test, Humans, Electronics, Medical
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 5 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |