doi: 10.1007/bf02133677
pmid: 3884455
The optics of slit-lamp fluorophotometry were analyzed. This analysis forms the basis of the formulation of a slit-lamp technique called photometric oculometry, which makes it possible to measure intraocular distances. A comparison between ultrasonography and photometric oculometry in the determination of the ocular axial length was carried out. Measurements in 11 eyes showed statistical agreement, with a mean of 22.90 mm, SD +/- 1.815 (ultrasound), and a mean of 23.17 mm, SD +/- 1.667 (photometry). The principle involved in photometric oculometry is of special importance for the calculation of the blood-retinal barrier permeability to fluorescein by vitreous fluorophotometry, since the method allows the volume of the vitreous body to be estimated directly from the slit-lamp fluorophotometer recording.
Photometry, Humans, Eye, Ultrasonography
Photometry, Humans, Eye, Ultrasonography
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 2 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |