doi: 10.1007/bf02035490
This paper discusses a novel application of PIGE for the determination of13C in breath. Samples of human breath, urea, glucose, benzamide, barium carbonate were analyzed against cylinder CO2 and graphite standard. An accuracy check of the13C determination (with reference to mass spectrometric “True” results) gave a relative error of only −0.4% for PIGE. The performance of different standards in this determination was assessed. Relative standard deviation for the determination of13C isotopic abundance in breath samples were <20%. Then, if a 25% change is conservatively assumed observable in13C abundance, an increase in13C percent isotopic abundance from the natural 1.11% (average) to only 1.39% may be detected.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 2 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |