doi: 10.1007/bf01983614
A round robin test was performed to determine the reliability of values for the glass transition temperatureTg as determined by DTA on polymers. Ten different instruments were involved. The test material was high molecular weight polystyrene. Values forTg (midpoint) were reported in the range 107°C±2 K. The respective heat flow curves differed considerably in shape. In the literature aTg of 100°C is often given for polystyrene. The discrepancy between this value and the value of 107°C found in the round robin test is due to three differences: the thermal history of the sample, the evaluation of the heat flow curves, and the effect of finite sample size.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 211 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 1% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 1% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |