doi: 10.1007/bf01979203
The reproducibility and reliability of the TA Instruments Modulated Differential Scanning Calorimeter (MDSC) was tested over a range of conditions. The equipment base line was found to be fairly constant with a very small fluctuation (10 μW), which means a 0.1 % fluctuation on the scale of a normal polymer MDSC curve. The excellent stability of the base line and the reasonable reproducibility of the curves (5%) suggest that frequent calibration is not required.