doi: 10.1007/bf00252168
pmid: 1031367
A total of 34 99mTc-generators from eight different manufactures were studied in 1974 and 1975 with regard to their performance and to the quality of the eluates. With some exceptions the generators performed well and proved their ability to act as a reliable source of 99mTc-pertechnetate. Two cases of mechanical malfunction and two cases with a low elution yield of 99mTc were observed. Four generators were overloaded with 99Mo. The 99mTc-pertechnetate did in most cases fulfil the requirements of the Nordic Pharmacopeia.
Radiochemistry, Evaluation Studies as Topic, Chemical Industry, Denmark, Technetium, Technology, Radiologic
Radiochemistry, Evaluation Studies as Topic, Chemical Industry, Denmark, Technetium, Technology, Radiologic
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 18 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 1% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |