AbstractPurposeZero‐echo‐time (ZTE) sequences have proven a powerful tool for MRI of ultrashort tissues, but they fail to produce useful images in the presence of strong field inhomogeneities (14 000 ppm). Here we seek a method to correct reconstruction artifacts from non‐Cartesian acquisitions in highly inhomogeneous , where the standard double‐shot gradient‐echo approach to field mapping fails.MethodsWe present a technique based on magnetic field maps obtained from two geometric distortion‐free point‐wise (SPRITE) acquisitions. To this end, we employ three scanners with varying field homogeneities. These maps are used for model‐based image reconstruction with iterative algebraic techniques (ART). For comparison, the same prior information is fed also to widely used Conjugate Phase (CP) algorithms.ResultsDistortions and artifacts coming from severe inhomogeneities, at the level of the encoding gradient, are largely reverted by our method, as opposed to CP reconstructions. This holds even close to the limit where intra‐voxel bandwidths (determined by inhomogeneities, up to 1.2 kHz) are comparable to the encoding inter‐voxel bandwidth (determined by the gradient fields, 625 Hz in this work).ConclusionWe have benchmarked the performance of a new method for ZTE imaging in highly inhomogeneous magnetic fields. For example, this can be exploited for dental imaging in affordable low‐field MRI systems, and can be expanded for arbitrary pulse sequences and extreme magnet geometries, as in, for example, single‐sided MRI.
Medical Physics, Phantoms, Imaging, Field inhomogeneity, Brain, Reproducibility of Results, FOS: Physical sciences, Physics - Medical Physics, Imaging Methodology, Magnetic Resonance Imaging, Magnetic Fields, Low field, Ultra‐short T2∗$$ {T}_2^{\ast } $$, Zero echo time, Image Processing, Computer-Assisted, Humans, Medical Physics (physics.med-ph), Artifacts, Artifact correction, Algorithms, Ultra-short T-2*
Medical Physics, Phantoms, Imaging, Field inhomogeneity, Brain, Reproducibility of Results, FOS: Physical sciences, Physics - Medical Physics, Imaging Methodology, Magnetic Resonance Imaging, Magnetic Fields, Low field, Ultra‐short T2∗$$ {T}_2^{\ast } $$, Zero echo time, Image Processing, Computer-Assisted, Humans, Medical Physics (physics.med-ph), Artifacts, Artifact correction, Algorithms, Ultra-short T-2*
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |