PurposeRespiratory motion makes hepatic ablation using high intensity focused ultrasound (HIFO) challenging. Previous HIFU liver treatment had required apnea induced during general anesthesia. We describe and test a system that allows treatment of the liver in the presence of breathing motion.MethodsMapping a signal from an external respiratory bellow to treatment locations within the liver allows the ultrasound transducer to be steered in real time to the target location. Using a moving phantom, three metrics were used to compare static, steered, and unsteered sonications: the area of sonications once a temperature rise of 15°C was achieved, the energy deposition required to reach that temperature, and the average rate of temperature rise during the first 10 s of sonication. Steered HIFU in vivo ablations of the porcine liver were also performed and compared to breath‐hold ablations.ResultsFor the last phantom metric, all groups were found to be statistically significantly different (P ≤ 0.003). However, in the other two metrics, the static and unsteered sonications were not statistically different (P > 0.9999). Steered in vivo HIFU ablations were not statistically significantly different from ablations during breath‐holding.ConclusionsA system for performing HIFU steering during ablation of the liver with breathing motion is presented and shown to achieve results equivalent to ablation performed with breath‐holding. Magn Reson Med 71:797–806, 2014. © 2012 Wiley Periodicals, Inc.
Respiratory-Gated Imaging Techniques, Phantoms, Imaging, Swine, Equipment Design, Magnetic Resonance Imaging, Equipment Failure Analysis, Treatment Outcome, Liver, Surgery, Computer-Assisted, Respiratory Mechanics, Animals, Hepatectomy, High-Intensity Focused Ultrasound Ablation, Artifacts
Respiratory-Gated Imaging Techniques, Phantoms, Imaging, Swine, Equipment Design, Magnetic Resonance Imaging, Equipment Failure Analysis, Treatment Outcome, Liver, Surgery, Computer-Assisted, Respiratory Mechanics, Animals, Hepatectomy, High-Intensity Focused Ultrasound Ablation, Artifacts
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 49 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |