pmid: 7476073
AbstractSpiral scanning gradient waveforms can be optimized with respect to blurring from off‐resonance effects by minimizing the readout time. This is achieved by maximizing the gradient amplitude during the scan so that the edge of k‐space is reached as quickly as possible. Gradient hardware constraints are incorporated by considering a circuit model for the gradient coil and amplifier. The optimized gradient waveforms are determined by a set of coupled differential equations. The resulting solutions have shorter readout time than solutions that do not consider the circuit model.
Models, Structural, Time Factors, Amplifiers, Electronic, Acceleration, Equipment Design, Image Enhancement, Magnetic Resonance Imaging, Algorithms
Models, Structural, Time Factors, Amplifiers, Electronic, Acceleration, Equipment Design, Image Enhancement, Magnetic Resonance Imaging, Algorithms
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 87 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 1% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |