doi: 10.1002/jor.25012
pmid: 33620100
AbstractThe sacroiliac joint (SIJ) is a known pain generator that, in severe cases, may require surgical fixation to reduce intra‐articular displacements and allow for arthrodesis. The objective of this computational study was to analyze how the number of implants affected SIJ stabilization with patient‐specific characteristics such as the pelvic geometry and bone quality. Detailed finite element models were developed to account for three pelvises of differing anatomy. Each model was tested with a normal and low bone density (LD) under two types of loading: compression only and compression with flexion and extension moments. These models were instrumented with one to three cylindrical, threaded and fenestrated implants through a posterior oblique trajectory, requiring less muscle dissection than the more common lateral trajectory used with triangular implants. Compared with the noninstrumented pelvis, the change in range of motion (ROM) and stress distribution were used to characterize joint stabilization. Noninstrumented mobility ranged from 0.86 to 2.55 mm and from 1.37° to 6.11°. Across patient‐specific characteristics, the ROM reduction with one implant varied from 3% to 21% for vertical and 15% to 47% for angular displacements. With two implants, the ROM reduction ranged from 12% to 41% for vertical and from 28% to 61% for angular displacements. Three implants, however, did not further improve the joint stability (14% to 42% for vertical and 32% to 63% for angular displacements). With respect to patient characteristics, an LD led to a decreased stabilization and a higher volume of stressed bone (>75% of yield stress). A better understanding of how patient characteristics affect the implant performance could help improve surgical planning of sacroiliac arthrodesis.
Spinal Fusion, Finite Element Analysis, [PHYS.MECA.BIOM] Physics [physics]/Mechanics [physics]/Biomechanics [physics.med-ph], Humans, Sacroiliac Joint, Prostheses and Implants, Range of Motion, Articular, Biomechanical Phenomena
Spinal Fusion, Finite Element Analysis, [PHYS.MECA.BIOM] Physics [physics]/Mechanics [physics]/Biomechanics [physics.med-ph], Humans, Sacroiliac Joint, Prostheses and Implants, Range of Motion, Articular, Biomechanical Phenomena
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 6 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |