doi: 10.1002/jms.5030
pmid: 38726700
AbstractWe consider the operation of a digital linear ion trap with resonant radial ejection. A sequence of rectangular voltage pulses with a dipole resonance signal is applied to the trap electrodes. The periodic waveform is piecewise constant, has zero mean, and is determined by an asymmetry parameter : one value is taken on interval and another on , where is the RF period. Ion mass scanning is performed by varying the asymmetry parameter and amplitude of the negative pulse part with time. The ion oscillation frequencies and acceptance of the linear trap are calculated. The dependence of the ion mass to charge ratio on the parameter is . The maximum value is about kDa for typical parameters of the linear trap: frequency 0.5 MHz, rod radius 4 mm, and negative pulse amplitude 1 kV. The dipolar excitation frequency is 0.125 MHz at which the LIT acceptance is maximal.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 2 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |