pmid: 7587733
AbstractBiohazardous aerosols generated during cell sorting have been of increased concern recently because of interest in sorting specimens containing human immunodeficiency virus type 1 (HIV‐1). Current flow cytometers have features designed to contain such aerosols within the sorting chamber, but the efficacy of these features has not been established. Therefore, we tested aerosol containment by two ELITE flow cytometers (Coulter Cytometry, Inc., Hialeah, FL) during sorting of specimens containing high titers of bacteriophage. Agar plates confluent with susceptible Escherichia coli were used to detect infectious units released from the sorting chamber. Under recommended operating conditions very few infectious units were released from the sorting chambers. Release increased when the center stream was not optimally collected in a vacuum‐exhausted tube or the chamber door was not completely closed. Failure of the negative pressure and high efficiency particle air (HEPA) filtration features had less of an effect. The data indicate that these standard safety features provide a rational expectation of safety for the flow cytometry operator. © 1995 Wiley‐Liss, Inc.
Aerosols, Evaluation Studies as Topic, HIV-1, Humans, Cell Separation, Containment of Biohazards, Flow Cytometry
Aerosols, Evaluation Studies as Topic, HIV-1, Humans, Cell Separation, Containment of Biohazards, Flow Cytometry
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 15 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |