pmid: 18623098
AbstractLeukocyte filters are applied clinically to remove leukocytes from blood. In order to optimize leukocyte filters, a mathematical model to describe the leukocyte filtration process was developed by modification of a general theoretical model for depth filtration. The model presented here can be used to predict the time‐dependent leukocyte filtration as a function of cell–cell interaction in the filter, filter efficiency, filter capacity, filter dimensions, and leukocyte concentration in the suspension applied to the filter. The results of different leukocyte filtration experiments previously reported in the literature could be well described by the present model. © 1995 John Wiley & Sons, Inc.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 3 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |