pmid: 27106664
We report a case demonstrating falsely elevated cyclosporine A (CsA) levels by using the antibody-conjugated magnetic immunoassay (ACMIA). To determine the cause of the false positive result, we examined the presence of an interferent in the patient's sample.A 26-year-old male patient displayed a high CsA concentration eight weeks after discontinuation of CsA treatment when measured by ACMIA. The possibility of assay interference was examined using serial dilution tests, heterophilic blocking tubes, erythrocyte washing, polyethylene glycol (PEG) protein precipitation, and a Protein A sepharose bead immunoabsorption assay.The serial dilution test showed nonlinearity, suggesting the presence of an interferent. CsA concentrations in the patient's plasma pre- and post-heterophilic blocking tube treatment were similar. Erythrocyte washing and PEG protein precipitation tests indicated the existence of a protein interferent in the plasma. Moreover, the Protein A bead immunoabsorption assay identified the interferent as antibodies against a unique epitope on the monoclonal antibody-β-galactosidase (mAb-β-gal) enzyme conjugate used in CsA ACMIA.We identified a circulating antibody against the unique epitope of the CsA ACMIA mAb-β-gal conjugate that induces falsely high CsA concentrations. These findings suggest that clinically aberrant CsA levels require immediate re-measurement by reference methods such as mass spectrometry.
Monoclonal/blood*, Adult, Immunoassay, Male, Magnetic Phenomena*, Magnetic Phenomena, Monoclonal/immunology*, 610, Antibodies, Monoclonal, Immunoassay*, Therapeutic drug monitoring, Antibodies, Epitopes/immunology*, Epitopes, Cyclosporine, Humans, Cyclosporine/blood*, Antibody-conjugated magnetic immunoassay, Assay interference
Monoclonal/blood*, Adult, Immunoassay, Male, Magnetic Phenomena*, Magnetic Phenomena, Monoclonal/immunology*, 610, Antibodies, Monoclonal, Immunoassay*, Therapeutic drug monitoring, Antibodies, Epitopes/immunology*, Epitopes, Cyclosporine, Humans, Cyclosporine/blood*, Antibody-conjugated magnetic immunoassay, Assay interference
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 6 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |