doi: 10.1364/oe.518821
pmid: 38859401
X-ray dual-phase grating interferometry provides quantitative micro-structural information beyond the optical resolution through its tunable correlation length. Ensuring optimal performance of the set-up requires accurate correlation length estimation and precise alignment of the gratings. This paper presents an automated procedure for determining the complete geometrical parameters of the interferometer set-up with a high degree of precision. The algorithm’s effectiveness is then evaluated through a series of experimental tests, illustrating its accuracy and robustness.
Technology and Engineering, Physics and Astronomy, FRINGE VISIBILITY, RAY
Technology and Engineering, Physics and Astronomy, FRINGE VISIBILITY, RAY
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |