The accuracy of gamma-spectrometric measurements in the isotopic analysis of shielded nuclear materials depends on the accuracy of the detector calibration in terms of energy efficiency, which should take into account corrections related to the measurement geometry and the absorption of γ-radiation by the shield material. The results of experimental studies of the energy efficiency of the HPGe-detector measured at fixed distances between the calibration source and the detector (50 and 100 mm) in the presence and absence of an absorbing shield made of stainless steel 12X18H10T (thickness ‒ 9.6 mm) are presented. On the basis of experimental data, an empirical description of the efficiency dependence for fixed distances between the gamma radiation source and the detector in the presence of a stainless steel absorbing shield was obtained.
Detectors and nuclear radiation detection
Detectors and nuclear radiation detection
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 2 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |