This dataset contains files of a reverse-phase HPLC-HRMS orbitrap measurement run during which the pump pressure gradually got more unstable. While a run was being recorded, the pump for the organic eluent ceased functioning slowly. This resulted in a high amount of new signals appearing in the form of "stripes" at specific retention times. It is intended as a control group for assigned data quality scores, with the reasoning that a highly unstable pressure in the chromatography will lead to more purely noise-induced signals. The relevant pressure profiles are also included. The mzML files were produced from the raw data using MSconvert and only contain the MS1 spectra. The last file in the series (20240805_AA_DK_Ibu_pH6_1min_20240808161549) remains incomplete since the instrument shut off halfway through the measurement.
Mass spectrometry, non-target screening, nontarget
Mass spectrometry, non-target screening, nontarget
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |