research product . Collection . 2006

Instrument for measuring the misalignments of ocular surfaces

Tabernero, Juan; Benito, Antonio; Nourrit, Vincent; Artal, Pablo;
  • Published: 01 Jan 2006
  • Publisher: Figshare
Abstract
A compact and robust instrument for measuring the alignment of ocular surfaces has been designed and used in living eyes. It is based on recording Purkinje images (reflections of light at the ocular surfaces) at nine different angular fixations. A complete analysis on the causes of misalignments of Purkinje images and its relations with those physical variables to be measured (global eye tilt, lens decentration and lens tilt) is presented. A research prototype based on these ideas was built and tested in normal and pseudophakic eyes (after cataract surgery). The new analysis techniques, together with the semicircular extended source and multiple fixation tests t...
Subjects
Medical Subject Headings: genetic structuressense organseye diseases