Abstract
This study is devoted to the estimation of effective dose for children undergoing interventional cardiology examinations. The conversion coefficients (CC) from directly measured dose area product (DAP) value to effective dose (ED) were calculated within the approved effective dose assessment methodology (Guidelines 2.6.1. 2944-11). The CC, Ed K , [mSv / (Gy • cm 2 )] for newborn infants and children of 1, 5, 10 and 15 years old (main(range)) were calculated as 2.5 (1.8-3.2); 1.1 (0.8-1.3); 0.6 (0.4-0.7); 0.4 (0.3-0.5); and 0,22 (0,18-0,30) respectively. A special Finnish computer program PCXMC 2.0 was used for calculating the dose CC. The series of calculations ...
Persistent Identifiers
Subjects
free text keywords: interventional cardiology, effective dose, pediatric patient, program pcxmc 2.0, lcsh:Medical physics. Medical radiology. Nuclear medicine, lcsh:R895-920, lcsh:Radioactivity and radioactive substances, lcsh:QC794.95-798, Pediatric age, Effective dose (radiation), Dose area product, medicine.medical_specialty, medicine, Age groups, Pediatrics, Conversion coefficients, business.industry, business, Cardiac interventions