A new measurement system enables combination of continuous glucose monitoring (CGM) and insulin infusion. A sensor system comprising an optical glucose biosensor and an optical oxygen sensor is integrated into the insulin infusion catheter of an insulin pump. Both sensors rely on near infrared (NIR) phosphorescent porphyrin dyes, wherefore the signals can be read out transcutaneous and non-invasively with a custom-built phase fluorometer measurement module. The spectral properties of the indicator dyes and the optical setup of the measurement module were optimized to enable independent read-out in two channels. Dynamic ranges from 0 mmHg to 160 mmHg oxygen and 0mg/dL to 360 mg/dL glucose (LOD 2mg/dL) are covered by the oxygen and the glucose sensor, respectively. In-vivo measurements in pigs demonstrate good correlation of reference blood glucose levels and glucose values obtained with the presented sensor system. The evaluation of the clinical accuracy of the system with Clarke Error Grid Analysis showed similar results to CGM-devices currently on the market.
Blood Glucose, Catheters, Swine, Blood Glucose Self-Monitoring, Biosensing Techniques, Equipment Design, Administration, Cutaneous, Cell Line, Insulin Infusion Systems, Animals, Humans, Insulin
Blood Glucose, Catheters, Swine, Blood Glucose Self-Monitoring, Biosensing Techniques, Equipment Design, Administration, Cutaneous, Cell Line, Insulin Infusion Systems, Animals, Humans, Insulin
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 40 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |