Simulations of a micro-PET System based on Liquid Xenon

Preprint English OPEN
Miceli, A. ; Glister, J. ; Andreyev, A. ; Bryman, D. ; Kurchaninov, L. ; Lu, P. ; Muennich, A. ; Retiere, F. ; Sossi, V. (2011)

The imaging performance of a high-resolution preclinical microPET system employing liquid xenon as the gamma ray detection medium was simulated. The arrangement comprises a ring of detectors consisting of trapezoidal LXe time projection ionization chambers and two array... View more
  • References (28)
    28 references, page 1 of 3

    [1] M. Bergeron, J. Cadorette, J.-F. Beaudoin, M.D. Lepage, G. Robert, V. Selivanov, M.-A. Tetrault, N. Viscogliosi, J.P. Norenberg, R. Fontaine, and R. Lecomte. Performance Evaluation of the LabPET APD-Based Digital PET Scanner. Nuclear Science, IEEE Transactions on, 56(1):10 -16, 2009.

    [2] Y. Tai, A. F. Chatziioannou, Y. Yang, R. W. Silverman, K. Meadors, S. Siegel, D. F. Newport, J. R. Stickel, and S. R. Cherry. MicroPET II: design, development and initial performance of an improved microPET scanner for small-animal imaging. Phys. Med. Biol., 48(11):1519-1537, 2003.

    [3] P. Sempere Roldan, E. Chereul, O. Dietzel, L. Magnier, C. Pautrot, L. Rbah, D. Sappey-Marinier, A. Wagner, L. Zimmer, M. Janier, V. Tarazona, and G. Dietzel. Raytest ClearPET(TM), a new generation small animal PET scanner. Nucl. Instr. and Meth. A, 571(1-2):498 - 501, 2007.

    [4] J. Seidel, J.J. Vaquero, J. Pascau, M. Desco, C.A. Johnson, and M.V. Green. Features of the NIH atlas small animal PET scanner and its use with a coaxial small animal volume CT scanner. Proceedings, IEEE International Symposium on Biomedical Imaging, page 545 548, 2002.

    [5] Yang Y, Wu Y, Qi J, St James S, Du H, Dokhale PA, Shah KS, Farrell R, and Cherry SR. A Prototype PET Scanner with DOI-Encoding Detectors. J Nucl Med, 49(7):1132 - 1140, 2008.

    [6] J. N. Marx et al. The Time Projection Chamber. Phys. Today, 31N10:46, 1978.

    [7] V. Chepel et al. Purification of Liquid Xenon and Impurity Monitoring for a PET Detector. Nucl. Instr. and Meth. A, 349:500-505, 1994.

    [8] E. Conti et al. Correlated Fluctuations between Luminescence and Ionization in Liquid Xenon. Phys. Rev. B, 68:054201, 2003.

    [9] M. Moszynski, M. Szawlowski, M. Kapusta, and M. Balcerzyk. Large area avalanche photodiodes in scintillation and X-rays detection. Nucl. Instr. and Meth. A, 485(3):504 - 521, 2002.

    [10] E. Aprile et al. Observation of Anti-correlation between Scintillation and Ionization for MeV Gamma-Rays in Liquid Xenon. Phys. Rev. B, Issue 1, 76:014115.

  • Similar Research Results (4)
  • Metrics
    No metrics available
Share - Bookmark