publication . Article . 2017

Treatment planning for MLC based robotic radiosurgery for brain metastases: plan comparison with circular fields and suggestions for planning strategies

Schmitt Daniela; El Shafie Rami; Klüter Sebastian; Arians Nathalie; Schubert Kai; Rieken Stefan; Debus Jürgen; Paul Angela;
Open Access
  • Published: 07 Sep 2017 Journal: Current Directions in Biomedical Engineering, volume 3, pages 151-154 (eissn: 2364-5504, Copyright policy)
  • Publisher: Walter de Gruyter GmbH
Abstract
<jats:title>Abstract</jats:title><jats:p>To evaluate the possible range of application of the new InCise2 MLC for the CyberKnife M6 system in brain radiosurgery, a plan comparison was made for 10 brain metastases sized between 1.5 and 9cm3 in 10 patients treated in a single fraction each. The target volumes consist of a PTV derived by expanding the GTV by 1mm and were chosen to have diversity in the cohort regarding regularity of shape, location and the structures needed to be blocked for beam transmission in the vicinity. For each case, two treatment plans were optimized: one using the MLC and one using the IRIS-collimator providing variable circular fields. Pl...
Subjects
free text keywords: Radiation treatment planning, Radiosurgery, medicine.medical_treatment, medicine, Medical physics, medicine.medical_specialty, business.industry, business, Treatment planning, stereotactic radiotherapy, robotic guided radiation therapy, multi leaf collimator, plan comparison, R
Powered by OpenAIRE Research Graph
Any information missing or wrong?Report an Issue