publication . Article . 2015

Test Method for High β Particle Emission Rate of 63Ni Source Plate

ZHANG Li-feng;
Open Access Chinese
  • Published: 01 Jan 2015 Journal: Journal of Isotopes (issn: 1000-7512, eissn: 1000-7512, Copyright policy)
  • Publisher: Editorial Board of Journal of Isotopes
Abstract
For the problem of measurement difficulties of β particle emission rate of Ni-63 source plate used for Ni-63 betavoltaic battery, a relative test method of scintillation current method was erected according to the measurement principle of scintillation detector.β particle emission rate of homemade Ni-63 source plate was tested by the method, and the test results were analysed and evaluated, it was initially thought that scintillation current method was a feasible way of testing β particle emission rate of Ni-63 source plate with high β particle emission rate.
Subjects
arXiv: Physics::Instrumentation and Detectors
free text keywords: Ni-63 betavoltaic battery Ni-63 source &beta particle emission rate, Nuclear and particle physics. Atomic energy. Radioactivity, QC770-798
Powered by OpenAIRE Research Graph