Hypoxemia in a patient with biventricular HeartWare®ventricular assist devices

Article English OPEN
Vélez-Martínez, Mariella ; Albaugh, Mary Ann ; Amin, Alpesh A. ; Thibodeau, Jennifer T. ; Mammen, Pradeep P. A. ; Markham, David W. ; Reimold, Sharon C. ; Drazner, Mark H. ; McKinnon, Leah J. ; Meyer, Dan M. ; Bethea, Brian T. ; Mishkin, Joseph D. (2013)
  • Publisher: Co-Action Publishing
  • Journal: Mechanical Circulatory Support
  • Related identifiers: doi: 10.3402/mcs.v4i0.22638
  • Subject: biventricular mechanical support; advanced heart failure; biventricular assist devices; HeartWare®; hypoxemia
    mesheuropmc: cardiovascular system | cardiovascular diseases | circulatory and respiratory physiology

We present the case of a 59-year-old Caucasian man with a nonischemic cardiomyopathy who underwent implantation of HeartWare® (HeartWare Inc., Framingham, MA) ventricular assist devices (HVAD) for biventricular support due to refractory cardiogenic shock after attempted ventricular tachycardia ablation. After implantation of biventricular HVADs, he developed hypoxemia that failed to correct with administration of supplemental oxygen. A patent foramen ovale (PFO) was discovered on postoperative transesophageal echocardiogram (TEE). Speed adjustment of both the right and left HVADs under TEE guidance allowed for spontaneous closure of the PFO and resolution of the hypoxemia.Keywords: biventricular mechanical support; advanced heart failure; biventricular assist devices; HeartWare®; hypoxemia(Published: 18 October 2013)Citation: Mechanical Circulatory Support 2013, 4: 22638 - http://dx.doi.org/10.3402/mcs.v4i0.22638
  • References (18)
    18 references, page 1 of 2

    1. Dell'Aquila AM, Schneider SR, Schlarb D, Redwan B, Sindermann JR, Ellger B, et al. Initial clinical experience with the HeartWare left ventricular assist system: a single-center report. Ann Thorac Surg 2013; 95(1): 170 7.

    2. Aaronson KD, Slaughter MS, Miller LW, McGee EC, Cotts WG, Acker MA, et al. Use of an intrapericardial, continuousflow, centrifugal pump in patients awaiting heart transplantation. Circulation 2012; 125(25): 3191 200.

    3. Krabatsch T, Potapov E, Stepanenko A, Schweiger M, Kukucka M, Huebler M, et al. Biventricular circulatory support with two miniaturized implantable assist devices. Circulation 2011; 124(Suppl 11): S179 86.

    4. Friedline K, Hassinger P. Total artificial heart freedom driver in a patient with end-stage biventricular heart failure. AANA J 2012; 80(2): 105 12.

    5. Loforte A, Stepanenko A, Potapov EV, Musumeci F, Dranishnikov N, Schweiger M, et al. Temporary right ventricular mechanical support in high-risk left ventricular assist device recipients versus permanent biventricular or total artificial heart support. Artif Organs. 2013; 37(6): 523 30.

    6. Patlolla B, Beygui R, Haddad F. Right-ventricular failure following left ventricle assist device implantation. Curr Opin Cardiol 2013; 28(2): 223 33.

    7. Kormos RL, Teuteberg JJ, Pagani FD, Russell SD, John R, Miller LW, et al. Right ventricular failure in patients with the HeartMate II continuous-flow left ventricular assist device: incidence, risk factors, and effect on outcomes. J Thorac Cardiovasc Surg 2010; 139(5): 1316 24.

    8. Hagen PT, Scholz DG, Edwards WD. Incidence and size of patent foramen ovale during the first 10 decades of life: an autopsy study of 965 normal hearts. Mayo Clin Proc 1984; 59(1): 17 20.

    9. Nakatani S, Thomas JD, Savage RM, Vargo RL, Smedira NG, McCarthy PM. Prediction of right ventricular dysfunction after left ventricular assist device implantation. Circulation 1996; 94(Suppl 9): II216 21.

    10. Marty P, Mejean S, Boudou N, Mayeur N, Minville V, Galinier M. Patent foramen ovale appearance with association of left ventricular assist device and mechanical ventilation. Am J Emerg Med 2012; 30(1): 259, e1 3.